Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Biological contamination

From Bioblast
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.


high-resolution terminology - matching measurements at high-resolution


Biological contamination

Description

Biological contamination may be caused by microbial growth in the O2k-Chamber or in the experimental medium.


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Biological contamination of the O2k-Chamber is detected as a high O2 flux in the closed chamber, measured in a QC1: Oxygen sensor test or QC2: Instrumental O2 background test.
The problem can be solved by first cleaning the O2k-Chamber by chemical sterilization (70 % ethanol with 30 % water; not 100 % ethanol) and test with a freshly prepared experimental medium, such as MiR05. Make sure the 70% ethanol does not contain any additives; e.g., 70 % ethanol used in hospital settings may contain antiseptics (EtOH containing antiseptics provided by Garcia-Roves PM). If repeated washing and storage with 70 % ethanol does not remove a biological contamination, the glass chamber has to be cleaned as described for protein contamination under O2k-Chamber cleaning with HCl by removing the glass chamber from the O2k and using a strong acid.