Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Open chamber"

From Bioblast
m
Line 1: Line 1:
{{Technical service}}
The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 [[flux at open chamber]].
The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 [[flux at open chamber]].
To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a [[stopper spacer]].
To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a [[stopper spacer]].
Line 10: Line 12:


{{#set:Technical service=Chamber| Technical service=POS calibration}}
{{#set:Technical service=Chamber| Technical service=POS calibration}}
[[Category:Technical service]]


__SHOWFACTBOX__
__SHOWFACTBOX__
{{Technical service}}

Revision as of 15:57, 13 February 2015

Template:Technical service

The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 flux at open chamber. To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a stopper spacer. This procedure is described in detail in the chapter "Air calibration" in [MiPNet12.08].

See also:

Air calibration

Closed chamber


... more about "Open chamber"
Chamber +  and POS calibration +