'''Problem:'''
Plotting of O2 flux per mass\ unit sample or flow per cell don't work - either via the manual selecting of the plot under menu 'Graph - Select Plots' or by choosing the appropriate layout under menu 'Layout'.
'''Cause:'''
# Sample Concentration\Amonut inside the 'Edit Experiment' window (menu 'Experiment - Edit') equals to 0
# The information of the Sample Unit\Concentration\Amonut inside the 'Edit Experiment' window (menu 'Experiment - Edit') hasn't been saved yet. Although the values inside this window are taken from the last experiment (measured with this DatLab installation, where the 'Save' button inside the 'Edit Experiment' has been pressed) and displayed at the begginig of a next experiments, they're not automatically saved.
'''Solution:'''
# Enter the Concentration\Amonut different from 0
# Inside the 'Edit Experiment' window (menu 'Experiment - Edit'), review (or enter) the sample information (Unit\Concentration\Amonut) click on the 'Save' button instead of 'Cancel'. Then (if the Concentration\Amonut not equals to 0) the appropriate plot can be selected.
[[User:Capek Ondrej|Capek Ondrej]] 15:24, 2 April 2015 (CEST)
{{#set:Technical service=DatLab}}
__SHOWFACTBOX__


[[Category:Technical service]]
[[Category:Technical service]]

Revision as of 15:24, 19 May 2015


                  


O2k-Open Support

Talk:DatLab plotting of O2 flux per mass\ unit sample or flow per cell don't work



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Problem: The plotting of the O2 flux per mass\unit sample or the flow per cell do not work - neither by selecting the plot manually in 'Graph - Select Plots' (in the menu) nor by choosing the appropriate layout from the menu 'Layout'.

Cause:

  1. Sample Concentration\Amount inside the 'Edit Experiment' window (menu 'Experiment - Edit') equal to 0
  2. The information of the Sample Unit\Concentration\Amount inside the 'Edit Experiment' window (menu 'Experiment - Edit') hasn't been saved yet. Although the values inside this window are taken from the last experiment (measured with this DatLab installation, where the 'Save' button inside the 'Edit Experiment' has been pressed) and displayed at the beginning of a next experiments, they are not automatically saved.

Solution:

  1. Enter the Concentration\Amount different from 0
  2. Inside the 'Edit Experiment' window (menu 'Experiment - Edit') review (or enter) the sample information (Unit\Concentration\Amount), click on the 'Save' button instead of 'Cancel' - then the appropriate plot can be selected (if the Concentration\Amount is not equal to 0).

Capek Ondrej 15:24, 2 April 2015 (CEST)