Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Volume-Calibration Ring"

From Bioblast
Line 1: Line 1:
{{Product
{{Technical support}}
|description='''Volume-Calibration Ring''' (A or B) PVDF for O2k-Stoppers (2-mL [[O2k-chamber]]), white; attached to the [[Stopper\black PEEK\conical Shaft\central Port |stoppers]].
__TOC__
|product id=2432A-01 or 2432B-01
{{MitoPedia without banner
|product type=O2k, Stopper
|abbr=
|description=[[Image:Volume calibration ring A+B PVDF for Stopper.jpg|right|180px]]'''Volume-Calibration Ring''' (A or B) PVDF for O2k-Stoppers (2-mL [[O2k-chamber]]), white; attached to the [[Stopper\black PEEK\conical Shaft\central Port |stoppers]].
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]]
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]]
|product image=[[Image:Volume calibration ring A+B PVDF for Stopper.jpg|right|180px]]
}}
}}
__TOC__
__TOC__
Line 17: Line 17:
== Keywords ==
== Keywords ==
{{Template:Keywords: Chamber volume}}
{{Template:Keywords: Chamber volume}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
}}

Revision as of 10:14, 21 October 2020


                  


O2k-Open Support

Volume-Calibration Ring



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Volume calibration ring A+B PVDF for Stopper.jpg

Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers (2-mL O2k-chamber), white; attached to the stoppers.


Reference: MiPNet22.11 O2k-FluoRespirometer manual



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Technical support

Volume calibration of the O2k-chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.