Volume-Calibration Ring

From Bioblast
Revision as of 10:53, 11 November 2020 by Tindle Lisa (talk | contribs)


                  


Oroboros Open Support

Volume-Calibration Ring



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: Oroboros Open Support 



Description

Volume calibration ring A+B PVDF for Stopper.jpg

Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers (2-mL O2k-chamber), white; attached to the stoppers.


Reference: Product details and purchase information


Technical support

Volume calibration of the O2k-chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.