Line 9: | Line 9: | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?''' | ||
::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental troubleshooting.-ยป[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]ยซ | ::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental troubleshooting.-ยป[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]ยซ | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?''' | ||
::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-ยป[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]ยซ | ::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-ยป[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]ยซ |
Revision as of 13:43, 10 March 2020
- Do you perform the leaky chamber test after chamber assembly?
- After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- ยปLeaky chamber testยซ
- Do you perform the leaky chamber test after chamber assembly?
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - ยปSlope smoothingยซ
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - ยปO2k-sV-Moduleยซ
- Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?
- Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental troubleshooting.-ยปSwitch components approachยซ
- Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?
- See how to clean and dry the electrical connections of the OroboPOS-Connector.-ยปCleaning the electrical connectionsยซ
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?