|
Β |
(31 intermediate revisions by 7 users not shown) |
Line 1: |
Line 1: |
| __TOC__
| |
| == While in home office mode ==
| |
| :::* [[User:Gnaigere|Gnaigere]] ([[User talk:Gnaigere|talk]]) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alertΒ -Β Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda '''analysis-linked''' O2k-Open Support cases during this time, and collect real-time operation cases for future releases.
| |
|
| |
|
|
| |
| == In discussion for page ==
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? '''
| |
| ::::::: - An unnoticed leak from the chamber may affect your data, therefore we recommend performing the ''Leaky chamber test'' - especially after chamber reassembly. - Β»[[O2k-Chamber_MitoPedia#Leaky_chamber_test|Leaky chamber test]]Β«
| |
| <br />
| |
|
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?'''
| |
| ::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform an experimental troubleshooting on your own.-Β»[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]Β«
| |
|
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?'''
| |
| ::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-Β»[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]Β«
| |
|
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]]Β [[File:Blue-info.png|40px|Info: improve]] '''Did you know that you can edit the background parameters in DatLab 7.4 O<sub>2</sub> analysis Excel template?'''
| |
| ::::::: If you forgot to apply the background correction before running your experiment, you can add the values directly to the O<sub>2</sub> analysis DL7.4 Excel template and complete your data analysis following the [[Oxygen flux - instrumental background|quality control step 2]]. Β»[[MiPNet24.06_Oxygen_flux_analysis_-_DatLab_7.4|Oxygen flux analysis - DatLab 7.4]]Β«
| |
|
| |
| == O2k-Open Support alert circulars ==
| |
|
| |
| === O2k-Open Support alert circular 2020(01): in prep ===
| |
| ::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.'''
| |
| ::::::: See how you can check that the two blue OroboPOS-Holders are intact. - Β»[[O2k-OroboPOS-Holder_MitoPedia#O2k-OroboPOS-Holder:_Leaky_chamber_due_to_damaged_OroboPOS-Holder |O2k-OroboPOS-Holder]]Β«
| |
| <br />
| |
| ::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|40px|Take care]] '''Calibration parameters are transferred to experimental DatLab files'''
| |
| ::::::: Save calibration information from selected [[Marks - DatLab |Marks]] while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - Β»[[Oxygen_calibration_-_DatLab#Oxygen_calibration_in_instrumental_and_experimental_DatLab_files | Oxygen calibration - DatLab]]Β«
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform O2k-Chamber volume calibration?'''
| |
| ::::::: The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the [[O2k-sV-Module]]). - Β»[[O2k-Chamber MitoPedia]]Β«
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you use the correct O<sub>2</sub> solubility factor (F<sub>M</sub>) in the βO2 calibrationβ window in DatLab?
| |
|
| |
| ::::::: The O<sub>2</sub> solubility factor varies according to the medium used and influences the air calibration. - Β»[[Oxygen solubility factor]]Β«
| |