Gnaiger E, Doeller JE, Kraus D, Shiva S, Brookes PS, Darley-Usmar VM. NO effect on mitochondrial oxygen kinetics at low oxygen. Oxygraph-2k workshop Report. Mitochondr Physiol Network 08.12.

Β» MiPNet08.12

OROBOROS (2003-2010) Mitochondr Physiol Network

Abstract: A single pilot experiment was carried out during an Oxygraph-2k workshop on high-resolution respirometry (IOC22). Respiration of isolated rat liver mitochondria was inhibited by addition of NO, which increased the sensitivity to oxygen >25-fold when compared to the half-saturation oxygen pressure, p50, in the absence of NO. Oxygen kinetics followed a monophasic hyperbolic function up to 2.2 kPa with NO (p50=0.93 kPa), compared to the standard oxygen range to 1.1 kPa without NO (p50=0.035 kPa).


β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS, US_AL Birmingham_Kraus DW


Labels: MiParea: Respiration, Instruments;methods 

Stress:Hypoxia, RONS; Oxidative Stress"RONS; Oxidative Stress" is not in the list (Cell death, Cryopreservation, Ischemia-reperfusion, Permeability transition, Oxidative stress;RONS, Temperature, Hypoxia, Mitochondrial disease) of allowed values for the "Stress" property.  Organism: Rat  Tissue;cell: Liver  Preparation: Isolated Mitochondria"Isolated Mitochondria" is not in the list (Intact organism, Intact organ, Permeabilized cells, Permeabilized tissue, Homogenate, Isolated mitochondria, SMP, Chloroplasts, Enzyme, Oxidase;biochemical oxidation, ...) of allowed values for the "Preparation" property. 

Regulation: Inhibitor, O2"O2" is not in the list (Aerobic glycolysis, ADP, ATP, ATP production, AMP, Calcium, Coupling efficiency;uncoupling, Cyt c, Flux control, Inhibitor, ...) of allowed values for the "Respiration and regulation" property.  Coupling state: OXPHOS 

HRR: Oxygraph-2k, NO, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 

O2k-Demo