O2k-Titration Set: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 16: | Line 16: | ||
== | == Handling of the syringes == | ||
* | * separate inhibitors from substrates | ||
* | * rinse the outside with water immediately after use | ||
* at the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent | |||
* | * rinse with pure solvent before use | ||
* between two runs during the day: rinse the outside with water | |||
* at the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100 % | |||
* storage: ethanol 100% (short), dry (longtime) | |||
Revision as of 09:18, 12 February 2013
O2k-Titration Set
Description | The Titration Set consists of 6x 10 Β΅l, 6x 25 Β΅l, and 2x 50 Β΅l Hamilton glass syringes, and 3 spare plungers for the 10 Β΅l syringes. It is a basic HRR-Accessory included in the O2k-Core Unit. |
---|---|
Product ID | 20820-01 |
Type | O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue |
Link | MiPNet09.12 O2k-Titrations |
Image |
The Titration Set consists of
- 6 Microsyringe\10 mm3 51/0.13 mm 10 mm3
- 6 Microsyringe\25 mm3 51/0.15 mm 25 mm3
- 2 Microsyringe\50 mm3 51/0.15 mm 50 mm3
- 3 Plunger\10 mm3 for 10 mm3 syringe
Handling of the syringes
- separate inhibitors from substrates
- rinse the outside with water immediately after use
- at the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent
- rinse with pure solvent before use
- between two runs during the day: rinse the outside with water
- at the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100 %
- storage: ethanol 100% (short), dry (longtime)