|description='''OroboPOS-Connector''' (blue POM), with male connection to [[OroboPOS]] sensor head (POS) and with cable and male plug fitting into [[O2k-Main Unit]].Two units of this item are standard components of the [[O2k-Assembly Kit]] ([[O2k-FluoRespirometer]]), mounted on the [[OroboPOS-Holder]]s at the [[O2k-Main Unit]].
|info=[[OroboPOS-Connector]]
}}
{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware
}}
{{MitoPedia topics}}
__TOC__
Β 
Β 
{{Template:Technical support integrated}}
Β 
Β 
== Functions ==
::: The OroboPOS-Connector has several different functions:
::::* It mechanically presses the [[POS]] against the [[glass chamber]], ensuring a tight fit.
::::* It connects the [[POS]] electronically to the [[Main unit]] transmitting the signal.
::::* In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the [[Main unit]]. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable.
Β 
:::: The OroboPOS-Connector is sensitive to damage by [[ESD]] when disconnected from the [[O2k-Main Unit]] and when the POS head is removed. Further details: [[MiPNet14.01_ESD-damage]].
Β 
== Cleaning the body of the OroboPOS-Connector ==
Β 
:::: When experimental media or electrolyte (during [[MiPNet19.18B POS-service |service of the polarographic oxygen sensor OroboPOS]]) leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water. To do so, screw the OroboPOS head onto the OroboPOS-Connector. Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part.
Β 
== Cleaning the electrical connections ==
Β 
::::# Unscrew the OroboPOS and check the electrical connections (the gold pin and thread on the OroboPOS and OroboPOS-Connector).
::::# Remove any contaminating grease and moisture with a fine paper cloth.
::::# Moisten a paper cloth with ethanol and carefully clean the gold pin and threads. If the gold pin and thread on the OroboPOS-Connector are discolored or appear stained, clean repeatedly with a paper towel moistened with ethanol, until no more traces appear on the paper towel used for cleaning.
::::* If the O2k is used in a humid environment, the connector can be dried in an oven at up to 60Β°C for 1-2h. Follow the procedure with an [[Oxygen sensor test]] as a quality control of proper OroboPOS-Connector and sensor performance.
Β 
== Replacing the OroboPOS-Connector ==
::::If you need to replace the Orobo-POS Connector, an [[OroboPOS-Connector Service]] may be possible.