Talk:Different O2 fluxes in left and right chamber: Difference between revisions

From Bioblast
No edit summary
(Blanked the page)
Line 1: Line 1:
{{Technical support}}


'''Question:''' In our experiments we get consistently higher fluxes in chamber A as compared to chamber B.
'''Answer:''' To exclude instrumental causes please follow the procedure described in the section
'''Calibration and quality control of the OroboPOS (O2k-SOP)''' in [[MiPNet06.03 POS-calibration-SOP]].
Possible instrumental and protocol related causes for fluxes differences between chambers:
* Incorrect O2 calibration in one chamber, see [[MiPNet06.03 POS-calibration-SOP]].
* Incorrect chamber volume calibration: The total amount of oxygen in the chamber with the smaller volume will be smaller. Therefore, the O2 concentration will drop faster resulting in a higher flux, see [[MiPNet03.02 Chemicals-Media| O2k-Series H: Start]].
* ##Incorrect background parameters: incorrect background parameters will cause an O2 concentration-dependent offset toΒ  the flux, the value of the offset will be independent of the absoluteΒ  value of the flux. E.g. a constant offset of 10 pmol / s ml at ca 100 Β΅MΒ  O2 concentration and 20 pmol/s ml at 300 Β΅M O2 concentration could be caused by wrong background parameters, butΒ  a constant difference of e.g.Β  20% at very different absolute values of the flux can not be caused byΒ  wrong background parameters. In any case, it is advisable to check theΒ  background parameters (perform an instrumental O2 background experiment). See [[MiPNet14.06 Instrumental O2 background]].##
* ##Sample injection: Scenario: One filling of a syringe is used to inject the sample into several chambers. Sedimentation starts to occur immediately and more sample will be injected into the first chamber than in anyΒ  subsequent chambers. A remedy we apply is to fill the syringe with sample sufficient for 3 chambers and inject the first aliquot back to the sample stock solution before quickly injecting the rest into the chambers. Independently of this we recommend randomizing the assignment of the chambers to prevent systematic errors.##
* Hydrophobic inhibitors, see [[MiPNet19.03 O2k-cleaning and ISS]].
* Biological contamination, see [[Biological contamination]].
HardwareΒ  defects: As hardware defects (sensors, electronics, etc.) have many negative effects (noisy signal, slow response, no signal, etc.) on experiments, it is difficult to see how a real hardware defect can cause a systematic error in flux calculation once a correct calibration of the O2 sensor (at air saturation and at zero oxygen) has been obtained. [[User:FaschingΒ  Mario|Fasching Mario]] 09:49, 7 November 2014 (CET)
{{#set: Technical service =Chamber}}
{{#set: Technical service =O2 signal}}
[[category:Technical service]]
__SHOWFACTBOX__

Revision as of 08:33, 24 April 2018

Cookies help us deliver our services. By using our services, you agree to our use of cookies.