== Leaky Chamber ==


=== Leaky chamber test ===
:::* When the leakage is so little that cannot be noticed in a short amount of time, we recommend to follow the ''Leaky chamber test'' below:
::::# Decrease the oxygen concentration down to ~ 50 ΞΌM (using [[Dithionite]] or N<sub>2</sub> injection)
::::# Let the O2k record overnight with the [[closed chamber|chamber closed]] and illumination switched off
::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure. In this case, the chamber is assembledΒ  properly.
::::# If the oxygen concentration increases overtime, a reassemble of the chamber is needed.
[[File:Leaking_chamber_2020-02-03 P2-03.png|1500px]]