Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:O2k-Chambers"

From Bioblast
(Created page with "== Leaky Chamber == === Leaky chamber test === :::* When the leakage is so little that cannot be noticed in a short amount of time, we recommend to follow the ''Leaky chamber...")
 
Line 7: Line 7:
::::# Let the O2k record overnight with the [[closed chamber|chamber closed]] and illumination switched off
::::# Let the O2k record overnight with the [[closed chamber|chamber closed]] and illumination switched off
::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure. In this case, the chamber is assembled  properly.
::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure. In this case, the chamber is assembled  properly.
::::# If the oxygen concentration increases overtime, a reassemble of the chamber is needed.
::::# If the oxygen concentration increases over time, a reassemble of the chamber is needed.
[[File:Leaking_chamber_2020-02-03 P2-03.png|1500px]]
[[File:Leaking_chamber_2020-02-03 P2-03.png|1500px]]
::::*'''Text in discussion for the page:'''
::::* To exclude the existence of an unnoticed leak from the chamber, we recommend to follow the ''Leaky chamber test'' below - especially after chamber reassembly:
::::# Decrease the oxygen concentration down to ~ 50 μM (using [[Dithionite]] or N<sub>2</sub> injection).
::::# Record the O2k measurement overnight with the [[closed chamber|chamber closed]] and illumination switched off.
::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure below. In this case, the chamber is assembled  properly.
::::# If the oxygen concentration increases over time, a reassemble of the chamber is needed.

Revision as of 15:51, 19 March 2020

Leaky Chamber

Leaky chamber test

  • When the leakage is so little that cannot be noticed in a short amount of time, we recommend to follow the Leaky chamber test below:
  1. Decrease the oxygen concentration down to ~ 50 μM (using Dithionite or N2 injection)
  2. Let the O2k record overnight with the chamber closed and illumination switched off
  3. The oxygen concentration should decrease overtime due to the O2 consumed by OroboPOS as it is shown in the figure. In this case, the chamber is assembled properly.
  4. If the oxygen concentration increases over time, a reassemble of the chamber is needed.

Leaking chamber 2020-02-03 P2-03.png


  • Text in discussion for the page:
  • To exclude the existence of an unnoticed leak from the chamber, we recommend to follow the Leaky chamber test below - especially after chamber reassembly:
  1. Decrease the oxygen concentration down to ~ 50 μM (using Dithionite or N2 injection).
  2. Record the O2k measurement overnight with the chamber closed and illumination switched off.
  3. The oxygen concentration should decrease overtime due to the O2 consumed by OroboPOS as it is shown in the figure below. In this case, the chamber is assembled properly.
  4. If the oxygen concentration increases over time, a reassemble of the chamber is needed.