{{Technical support integrated}}
'''Question:'''
'''Question:'''
I guess I would also like to know you guys opinion on TPP vs. safranin for measuring membrane potential.  Which is better? Which is more reliable?  Which is easier?
I guess I would also like to know you guys opinion on TPP vs. safranin for measuring membrane potential.  Which is better? Which is more reliable?  Which is easier?


 
'''Answer by OROBOROS:'''
'''Answer by OROBOROS:'''
'''Update:''' please see also: [[MiPNet19.19 Safranin Data Acquisition and Analysis]] and [[Krumschnabel 2014 Methods Enzymol]].


One of the big issues with measuring mitochondrial membrane potential is how to get "absolute values", please see
One of the big issues with measuring mitochondrial membrane potential is how to get "absolute values", see


[[Mitochondrial membrane potential]]
[[Mitochondrial membrane potential]]
Line 33: Line 33:
* The decision also relies on the types of calibration methods that may be required.
* The decision also relies on the types of calibration methods that may be required.
* There is probably always a benefit of having both methods available.
* There is probably always a benefit of having both methods available.
* [[Talk:Safranin#Inhibition_by_safranin|Inhibition by safranin]] may also be a major disadvantage of the safranin method.
* Inhibition by safranin is a problem to be considered with all cations used to measure mtMP.
 
 
'''More input from users to this important topic would be welcome. Please contact  instruments@oroboros.at to get an account for this wiki to contribute to this discussion'''!




'''Follow up question''': Whats the difference between a TPP and a safranin protocol?
'''Follow up question''': Whats the difference between a TPP and a safranin protocol?


'''Answer:''' We have to discern between the protocol for the addition of substrates, uncouplers, etc, referred to here as the [[Substrate-uncoupler-inhibitor titration|"SUIT  protocol"]] and specific additional steps necessary for safranin or TPP, referred to as the "specific protocol". Unlike other methods neither safranin nor TPP require major modifications to the "SUIT protocol", apart from excluding incompatible chemicals. That is to say, both methods can be used with a wide range of different SUIT protocols. The specific protocol for TPP comprises calibration of the TPP electrode, introduction of sample into the chamber containing TPP, etc. All this is done "around" and in addition to the SUIT protocol.
'''Answer:''' We have to discern between the [[Substrate-uncoupler-inhibitor titration|SUIT  protocol]] protocol for the addition of substrates, uncouplers, etc, and additional steps necessary for safranin or TPP. Neither safranin nor TPP require major modifications to the SUIT protocol, apart from excluding incompatible chemicals. Both methods can be used with a wide range of different SUIT protocols.
The "specific protocol" for safranin is quite straighforward. In the normal case of using it in a spectrofluorometer cuvette, this would include:
 
# Deciding on a safranin concentration and sample-to-safranin concentration ratio.
# Selection of excitation and emission wavelength and bandwidth.
# Addition of safranin.
# Addition of sample and start of SUIT protocol.
 
'''Update:''' please see also: [[MiPNet19.19 Safranin Data Acquisition and Analysis]] and [[Krumschnabel 2014 Methods Enzymol]].
 
Calibration of the safranin fluorescence by comparing the results with membrane potentials obtained by another method should be done in a separate experiment.
 
 
If using the [[O2k-Fluorescence LED2-Module]] the second step is replaced by selecting an appropriate sensor, filter set and light intensity, see [[O2k-Fluorescence_LED2-Module#Application_specific_settings|application specific settings]]. 
 
For an example of a protocol used together with safranin see [[Komary 2010 Biochim Biophys Acta]].
 
 


--[[User:Fasching Mario|Fasching Mario]] 12:32, 4 May 2012 (CEST)
--[[User:Fasching Mario|Fasching Mario]] 12:32, 4 May 2012 (CEST)


== Safranin calibration ==
== Safranin calibration ==


'''Question:'''  
'''Question:'''  
We've been using the O2K unit to measure ROS production using amplex ultra red, and so far I think it has been going well!   
We've been using the O2K unit to measure ROS production using amplex ultra red, and so far I think it has been going well!   
I now would like to use saffranin to measure membrane potential. I  just read up on some info on the O2K website. Do I have to calibrate the signal as I do with amplex red?  And if so, what should I use?  Is it  done in the same way (with the same excel spreadsheet?
I now would like to use safranin to measure membrane potential. I  just read up on some info on the O2K website. Do I have to calibrate the signal as I do with Amplex red?  And if so, what should I use?  Is it  done in the same way (with the same excel spreadsheet)?


'''Answer:'''
'''Answer:'''
* With DatLab 6 an Excel spreadsheet is no longer required: »[[MiPNet19.19 DatLab 6]]
* [[Krumschnabel 2014 Methods Enzymol]]
* [[MiPNet20.13 Safranin_mt-membranepotential]]
* [[Figueira 2012 Methods Mol Biol]]


'''Update:''' please see also: [[MiPNet19.19 Safranin Data Acquisition and Analysis]] and [[Krumschnabel 2014 Methods Enzymol]].
In many publications, e.g. [[Komary 2010 Biochim Biophys Acta]], no calibration was done and only the plot of the fluorescence  intensities is presented. A simple two point or multiple point calibration for safranin concentration is recommended, and a correction for chemical background effects. A claimed linear relationship with the mitochondrial membrane potential is based on the safranin fluorescence signal and not the actual safranin concentration. When safranin is titrated in the presence of sample, safranin is immediately taken up by the sample. To a certain degree this can be compensated for by placing a short mark immediately after the injection but the accuracy of this compensation will depend on the speed of uptake of safranin by the sample. [[User:Fasching Mario|Fasching Mario]] 09:13, 10 January 2013 (CET)
That depends on whether quantification is required and if so what kind. Safranin is mainly a  qualitative method, however for isolated mitochondria the literature claims that for certain safranin to sample concentration ratios plus well defined  absolute safranin concentrations there is linear correlation between  detected safranin fluorescence and mitochondrial membrane potential, e.g. [[Figueira 2012 Methods Mol Biol]].
 
For other samples (e.g. permeabilised cells) this relation, as shown in  the literature, is definitely non-linear. In the first instance, it is advisable to consult the literature published by workers using safranin in cuvettes to see what calibration procedures they used, and at what point they introduced the safranin. However, in many papers, e.g. [[Komary 2010 Biochim Biophys Acta]], no calibration was done and only the plot of the fluorescence  intensities is presented. This is an honest, pragmatic approach. Nonetheless, independent of your final format for publication, as a  minimum, a simply two point calibration for safranin concentration is recommended. This you can do directly in DatLab in the  calibration window. This should be enough if you mainly want to look on  the data in a qualitative way but will still allow you to compare  results from different sensors and do a correction for any chemical background effects.
 
If quantification is required, a multiple point calibration could be used to obtain more precise safranin concentrations. However the claimed linear relationship with the mitochondrial membrane potential is based on the safranin fluorescence signal and not the actual safranin concentration. If you want to do a multi point calibration you can indeed use the template for Amplex red. So when should the safranin be introduced into the chamber for calibration purposes? If you inject the safranin for the calibration  before the  sample it will be easy to set the markers. However, the very different absorption properties of the solution without the sample compared to that with the sample means that sensitivity will probably be quite different. When the safranin is injected after the sample, the sensitivity will be  correct but safranin is immediately taken up by the sample. To a certain degree this can be compensated for by placing a short mark immediately after the injection but the accuracy of this compensation will depend on the speed of uptake of safranin by the sample. However,  If full quantification is not required, any small error here should be negligible. If the display of fluorescence intensity in a plot together with plots from two different sensors is desired, this (displaying roughly calibrated safranin concentrations) could be a recommended approach, otherwise there will be an offset between traces from the two different sensors. Indeed, just for the purpose of normalization between different sensors even calibration without biological material would be suitable, maybe labelling the y axis of the resulting plot "arbitrary units" which is possible in the newest release of [[DatLab]] 5. [[User:Fasching Mario|Fasching Mario]] 09:13, 10 January 2013 (CET)
 
== Inhibition by safranin ==
 
Safranin inhibits mitochondrial function ([[Krumschnabel 2014 Methods Enzymol]]). Always check the influence of the safranin concentration used on the respiratory rate. Insufficient data is available to define "safe" safranin concentrations for different sample types. Please add your own experiences here. [[User:Fasching Mario|Fasching Mario]] 09:12, 10 January 2013 (CET)
 
This problem seems to be quite severe. [[User:Laner Verena|Laner Verena]] 14:36, 16 April 2013 (CEST)
 
== Safranin chemical background ==
 
Several substances typically used in SUIT protocols may influence the fluorescence signal from safranin when injected into the O2k-Chamber. The chemical used should be tested for this effect in a background run without a biological sample. If necessary corrections should be applied. Strongly colored substances such as cytochrome c can be expected to have such an effect. A significant effect has also been found with ADP. [[User:Fasching Mario|Fasching Mario]] 09:19, 10 January 2013 (CET)
 
 
'''Substances with an effect on the fluorescence signal of safranin'''
* ADP (D)
* Cytochrome ''c'' (c)
* Succinate (S)
* Rotenone (Rot)
* Ascorbate (As)
* TMPD (Tm)
 


'''Substances without an effect on the fluorescence signal of safranin'''


* Pyruvate (P)
[[Image:BB-Bioblast.jpg|left|40px|link=http://www.bioblast.at/index.php/Bioblast:About|Bioblast wiki]]
* Malate (M)
== Popular Bioblast page ==
* Glutamate (G)
[[Safranin]] has been accessed more than
* Digitonin (Dig)
::::* 10,000 times (2019-08-07)
* Oligomycin (Omy)
::::* 5,000 times (2016-02-27)
* FCCP (U)
* Malonic acid (Mna)
* Antimycin A (Ama)
* DMSO
* Ethanol
* H<sub>2</sub>O<sub>2</sub>
* H<sub>2</sub>O

Latest revision as of 15:14, 7 August 2019

TPP vs safranin


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Question: I guess I would also like to know you guys opinion on TPP vs. safranin for measuring membrane potential. Which is better? Which is more reliable? Which is easier?


Answer by OROBOROS:

One of the big issues with measuring mitochondrial membrane potential is how to get "absolute values", see

Mitochondrial membrane potential and especially

Calculation of mitochondrial membrane potential from measurements with a TPP electrode

Absolute measurements of membrane potential in isolated mitochondria can be obtained with the TPP method by using unspecific binding correction factors that have been obtained via a "calibration" against the Rb radio labeling method. For other sample types "absolute" quantification has still to be developed, see the link above.

The safranin method works totally differently. Unlike the TPP method the relationship between fluorescence and membrane potential is entirely empirical. A linear relationship between fluorescence intensity and mitochondrial membrane potential is found for certain ranges and ratios of safranin and mitochondrial concentrations. In order to obtain quantitative values for mitochondrial membrane potential from the safranin method two caveats apply:

  • It must be established that the relationship is linear (or at least known) for the experimental conditions.
  • Some calibration is necessary, even for an established linear relationship.

One method to calibrate the safranin method is actually to use the TPP method, though you will find other methods in the literature. That in fact was the driving force for some customers, who had already established the safranin technique method using spectrofluormetry, subsequently to obtain a TPP electrode as well.

On the the other hand the safranin method is far easier to handle in the lab than the TPP method (no extra electrodes, no problem with carry-over of inhibitors by the electrode, etc.). Safranin can be added to many standard protocols and the changes in fluorescence intensity (at least qualitatively) can be followed in parallel to the respiration measurements without going to all the extra steps necessary for TPP measurements.

Summarising, the differences between the safranin and TPP methods are:

  • If absolute measurements of membrane potential in isolated mitochondria are required, TPP is the method of choice.
  • If quantification of differences is sufficient then safranin and TPP are both suitable.
  • The type of sample may make one method or the other more suitable.
  • The decision also relies on the types of calibration methods that may be required.
  • There is probably always a benefit of having both methods available.
  • Inhibition by safranin is a problem to be considered with all cations used to measure mtMP.


Follow up question: Whats the difference between a TPP and a safranin protocol?

Answer: We have to discern between the SUIT protocol protocol for the addition of substrates, uncouplers, etc, and additional steps necessary for safranin or TPP. Neither safranin nor TPP require major modifications to the SUIT protocol, apart from excluding incompatible chemicals. Both methods can be used with a wide range of different SUIT protocols.

--Fasching Mario 12:32, 4 May 2012 (CEST)


Safranin calibration

Question: We've been using the O2K unit to measure ROS production using amplex ultra red, and so far I think it has been going well! I now would like to use safranin to measure membrane potential. I just read up on some info on the O2K website. Do I have to calibrate the signal as I do with Amplex red? And if so, what should I use? Is it done in the same way (with the same excel spreadsheet)?

Answer:

In many publications, e.g. Komary 2010 Biochim Biophys Acta, no calibration was done and only the plot of the fluorescence intensities is presented. A simple two point or multiple point calibration for safranin concentration is recommended, and a correction for chemical background effects. A claimed linear relationship with the mitochondrial membrane potential is based on the safranin fluorescence signal and not the actual safranin concentration. When safranin is titrated in the presence of sample, safranin is immediately taken up by the sample. To a certain degree this can be compensated for by placing a short mark immediately after the injection but the accuracy of this compensation will depend on the speed of uptake of safranin by the sample. Fasching Mario 09:13, 10 January 2013 (CET)