* Recommendation: Set "Slope smoothing" to 20 for performing and analyzing experiments with biological sample (strongly recommended in the case of uncoupler titrations). A slope smoothing of 20 increases the time or wavelength | * Recommendation: Set "Slope smoothing" of the Amp slope to 20 for performing and analyzing experiments with biological sample (strongly recommended in the case of uncoupler titrations). A slope smoothing of 20 increases the time or wavelength resolution, but increases the noise of the signal. | ||
:The following DatLab traces illustrate an example of "Slope smoothing" using 20 versus 40 data points: | :The following DatLab traces illustrate an example of "Slope smoothing" using 20 versus 40 data points: | ||
Latest revision as of 12:09, 4 March 2020
- Recommendation: Set "Slope smoothing" of the Amp slope to 20 for performing and analyzing experiments with biological sample (strongly recommended in the case of uncoupler titrations). A slope smoothing of 20 increases the time or wavelength resolution, but increases the noise of the signal.
- The following DatLab traces illustrate an example of "Slope smoothing" using 20 versus 40 data points:
- Slope smoothing: 20
- Slope smoothing: 40