Zero calibration is together with air calibration one of the two steps of the [[POS calibration]], see [[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]]. It is done at [[closed chamber]] after all the oxygen has been removed by the addition of [[dithionite]]. A zero calibration is the last part of the [[sensor test]] and described in detail there. Unlike [[air calibration]] it is not necessary to perform a zero calibration each day.
{{MitoPedia
|abbr=R1
|description='Zero calibration is together with [[air calibration]] one of the two steps of the [[POS calibration]]. It is done at [[closed chamber]] after all the oxygen has been removed by the addition of [[dithionite]]. A zero calibration is the last part of the [[sensor test]] and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.
|info=[[D MiPNet12.08 O2k-Calibration]]
}}
{{MitoPedia methods
|mitopedia method=Respirometry
}}
Β 
== Zero calibration in the O2k ==


'''A zero calibration should be performed''' Β 
'''A zero calibration should be performed''' Β 

Revision as of 11:26, 20 June 2013


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the POS calibration. It is done at closed chamber after all the oxygen has been removed by the addition of dithionite. A zero calibration is the last part of the sensor test and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R1

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 


Zero calibration in the O2k

A zero calibration should be performed

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used



Template:Technical service

... more about "Zero calibration"
'Zero calibration is together with air calibration … 'Zero calibration is together with air calibration one of the two steps of the POS calibration. It is done at closed chamber after all the oxygen has been removed by the addition of dithionite. A zero calibration is the last part of the sensor test and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +