Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Zero calibration"

From Bioblast
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|abbr=R1
|abbr=R0
|description='Zero calibration is together with [[air calibration]] one of the two  steps of the [[POS calibration]].  It is done at [[closed chamber]] after all the oxygen has been removed  by the addition of [[dithionite]]. A zero calibration is the last part  of the [[sensor test]] and described in detail there. Unlike air  calibration it is not necessary to perform a zero calibration each day.
|description='Zero calibration is together with [[air calibration]] one of the two  steps of the [[POS calibration]].  It is done at [[closed chamber]] after all the oxygen has been removed  by the addition of [[dithionite]]. A zero calibration is the last part  of the [[sensor test]] and described in detail there. Unlike air  calibration it is not necessary to perform a zero calibration each day.
|info=[[D MiPNet12.08 O2k-Calibration]]
|info=[[D MiPNet12.08 O2k-Calibration]]

Revision as of 15:05, 16 October 2013


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the POS calibration. It is done at closed chamber after all the oxygen has been removed by the addition of dithionite. A zero calibration is the last part of the sensor test and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R0

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 


Zero calibration in the O2k

A zero calibration should be performed

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used



Template:Technical service

... more about "Zero calibration"
'Zero calibration is together with air calibration'Zero calibration is together with air calibration one of the two steps of the POS calibration. It is done at closed chamber after all the oxygen has been removed by the addition of dithionite. A zero calibration is the last part of the sensor test and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +
Respirometry +
POS calibration +