Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Zero calibration"

From Bioblast
Line 1: Line 1:
{{Technical service}}
{{MitoPedia
{{MitoPedia
|abbr=R0
|abbr=R0
Line 21: Line 22:


{{#set:Technical service=POS calibration}}
{{#set:Technical service=POS calibration}}
[[Category:Technical service]]


__SHOWFACTBOX__
__SHOWFACTBOX__
{{Technical service}}

Revision as of 17:01, 13 February 2015

Template:Technical service


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R0

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 



Zero calibration in the O2k

A zero calibration should be performed

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used


... more about "Zero calibration"
'Zero calibration is together with air calibration … 'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +
Respirometry +
POS calibration +