Difference between revisions of "Zero calibration"
From Bioblast
Capek Ondrej (talk | contribs) |
|||
Line 1: | Line 1: | ||
{{Technical service}} | |||
{{MitoPedia | {{MitoPedia | ||
|abbr=R0 | |abbr=R0 | ||
Line 21: | Line 22: | ||
{{#set:Technical service=POS calibration}} | {{#set:Technical service=POS calibration}} | ||
[[Category:Technical service]] | |||
__SHOWFACTBOX__ | __SHOWFACTBOX__ | ||
Revision as of 17:01, 13 February 2015
Description
'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.
Abbreviation: R0
Reference: D MiPNet12.08 O2k-Calibration
MitoPedia methods:
Respirometry
Zero calibration in the O2k
A zero calibration should be performed
- from time to time as a record of instrumental performance: use the template O2k-Calibration-list.
A zero calibration has to be performed
- when a new or serviced POS is used for the first time
- when a new membrane was applied to the sensor
- when the gain is changed
- when a new POS connector is used
... more about "Zero calibration"
'Zero calibration is together with air calibration β¦ 'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +
Respirometry +
POS calibration +