Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Zero calibration"

From Bioblast
Line 1: Line 1:
{{Technical service}}
{{MitoPedia
{{MitoPedia
|abbr=R0
|abbr=R0
Line 9: Line 8:
}}
}}
{{MitoPedia topics}}
{{MitoPedia topics}}
== Zero calibration in the O2k ==


== Zero oxygen calibration in the O2k ==
{{Technical support integrated}}
'''A zero calibration should be performed''' Β 
'''A zero calibration should be performed''' Β 
* from time to time as a record of instrumental performance: use the template [http://www.oroboros.at/index.php?id=o2k-o2calibration O2k-Calibration-list].
* From time to time as a record of instrumental performance: use the template [http://www.oroboros.at/?o2k-o2calibration O2k-Calibration-list].


'''A zero calibration has to be performed'''
'''A zero calibration has to be performed'''


* when a new or serviced [[POS]] is used for the first time
* When measurements are performed at low oxygen (<10 Β΅M).
* when a new membrane was applied to the sensor
* Bracketing zero calibrations are required when measurements are performed near the limit of detection at oxygen concentrations close to 5 nM (0.005 Β΅M).
* when the [[gain (O2 channel)|gain]] is changed
* When a new or serviced [[OroboPOS]] is used for the first time.
* when a new [[POS connector]] is used
* When a new membrane was applied to the sensor.
* When the [[gain (O2 channel)|gain]] is changed.
* When a new [[POS connector]] is used.


{{#set:Technical service=POS calibration}}
{{#set:Technical service=POS calibration}}
[[Category:Technical service]]
[[Category:Technical service]]
__SHOWFACTBOX__

Revision as of 11:20, 16 May 2015


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R0

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 



Zero oxygen calibration in the O2k


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



A zero calibration should be performed

A zero calibration has to be performed

  • When measurements are performed at low oxygen (<10 Β΅M).
  • Bracketing zero calibrations are required when measurements are performed near the limit of detection at oxygen concentrations close to 5 nM (0.005 Β΅M).
  • When a new or serviced OroboPOS is used for the first time.
  • When a new membrane was applied to the sensor.
  • When the gain is changed.
  • When a new POS connector is used.