Zero calibration

From Bioblast


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R0

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 



Zero calibration in the O2k

A zero calibration should be performed

  • from time to time as a record of instrumental performance: use the template [[1]].

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used



Template:Technical service

... more about "Zero calibration"
'Zero calibration is together with air calibration … 'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +