Talk:O2k-Open Support alert: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
__TOC__ | __TOC__ | ||
== While in home office mode == | |||
:::* [[User:Gnaigere|Gnaigere]] ([[User talk:Gnaigere|talk]]) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alertΒ -Β Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda '''analysis-linked''' O2k-Open Support cases during this time, and collect real-time operation cases for future releases. | |||
== In discussion for page == | == In discussion for page == | ||
Revision as of 15:23, 19 March 2020
While in home office mode
- Gnaigere (talk) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alert - Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda analysis-linked O2k-Open Support cases during this time, and collect real-time operation cases for future releases.
In discussion for page
- Do you perform the leaky chamber test after chamber assembly?
- - An unnoticed leak from the chamber may affect your data, therefore we recommend performing the Leaky chamber test - especially after chamber reassembly. - Β»Leaky chamber testΒ«
- Do you perform the leaky chamber test after chamber assembly?
- Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?
- Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform an experimental troubleshooting on your own.-Β»Switch components approachΒ«
- Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?
- See how to clean and dry the electrical connections of the OroboPOS-Connector.-Β»Cleaning the electrical connectionsΒ«
- Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?
O2k-Open Support alert circulars
O2k-Open Support alert circular 2020(01): in prep
- New A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
- See how you can check that the two blue OroboPOS-Holders are intact. - Β»O2k-OroboPOS-HolderΒ«
- New A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
- New Calibration parameters are transferred to experimental DatLab files
- Save calibration information from selected Marks while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - Β» Oxygen calibration - DatLabΒ«
- Do you perform O2k-Chamber volume calibration?
- The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the O2k-sV-Module). - Β»O2k-Chamber MitoPediaΒ«
- Do you use the correct O2 solubility factor (FM) in the βO2 calibrationβ window in DatLab?
- New Calibration parameters are transferred to experimental DatLab files
- The O2 solubility factor varies according to the medium used and influences the air calibration. - Β»Oxygen solubility factorΒ«