Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "SUIT-009 O2 mt D015"

From Bioblast
Line 16: Line 16:
::: '''[[Categories of SUIT protocols |SUIT-category]]:''' NS(P)
::: '''[[Categories of SUIT protocols |SUIT-category]]:''' NS(P)
::: '''[[SUIT protocol pattern]]:''' orthogonal 1S;2D;3P;4Rot  
::: '''[[SUIT protocol pattern]]:''' orthogonal 1S;2D;3P;4Rot  
SUIT 9 is a short protocol for simultaneous determination of O<sub>2</sub> flux and the rate of [[MiPNet20.14 AmplexRed H2O2-production|H<sub>2</sub>O<sub>2</sub> production]] (H<sub>2</sub>O<sub>2</sub> flux) on isolated mitochondria,tissue homogenate (except of liver) and permeabilized cells. Succinate (S) supports the reverse electron transfer (RET) initiated H<sub>2</sub>O<sub>2</sub> production in the LEAK state. Addition of ADP initiates S-OXPHOS capacity. NS pathway control is induced by addition of succinate and [[NADH Electron transfer-pathway state|NADH-linked substrates]] such as Pyruvate (P).  
Originially this is a short protocol for simultaneous determination of O<sub>2</sub> flux and the rate of [[MiPNet20.14 AmplexRed H2O2-production|H<sub>2</sub>O<sub>2</sub> production]] (H<sub>2</sub>O<sub>2</sub> flux) on isolated mitochondria,tissue homogenate (except of liver) and permeabilized cells. It is used as a control experiment for[[SUIT-009 AmR mt D021]] without using the Amplex UltraRed assay in order to detect the effect of the fluorescence dye on the O<sub>2</sub> flux. We do not recommend this protocol for determination of the O<sub>2</sub> flux in the absence of Rotenone.  
For O2-Application, choose [[SUIT-009 O2 mt D015]]; for O2 and H<sub>2</sub>O<sub>2</sub> measurements (AmR), choose: D021_1S;2D;3P;4Rot;5Ama.
For O2-Application with permeabiluzed cells, choose [[SUIT-009 O2 pce D016]]; for O2 and H<sub>2</sub>O<sub>2</sub> measurements (AmR), choose: [[SUIT-009 AmR mt D021]].


__TOC__
__TOC__

Revision as of 11:11, 28 January 2019


high-resolution terminology - matching measurements at high-resolution


SUIT-009 O2 mt D015

Description

SUIT9

Abbreviation: NS(P)

Reference: A: SUIT-009 short protocol for respiration in mitochondrial preparations (isolated mitochondria and tissue homogenate).

SUIT number: D015_1S;2D;3P;4Rot;5Ama


MitoPedia concepts: SUIT protocol, SUIT B, Find 


MitoPedia methods: Respirometry 




SUIT-category: NS(P)
SUIT protocol pattern: orthogonal 1S;2D;3P;4Rot

Originially this is a short protocol for simultaneous determination of O2 flux and the rate of H2O2 production (H2O2 flux) on isolated mitochondria,tissue homogenate (except of liver) and permeabilized cells. It is used as a control experiment forSUIT-009 AmR mt D021 without using the Amplex UltraRed assay in order to detect the effect of the fluorescence dye on the O2 flux. We do not recommend this protocol for determination of the O2 flux in the absence of Rotenone. For O2-Application with permeabiluzed cells, choose SUIT-009 O2 pce D016; for O2 and H2O2 measurements (AmR), choose: SUIT-009 AmR mt D021.

Communicated by Iglesias-Gonzalez J,  Komlodi T and Gnaiger E (last update 2019-01-22)

DLP applications

References

MitoPedia: SUIT

Steps and respiratory states

1S;2D;3P;4Rot;5Ama.png

Step State Pathway Q-junction Comment - Events (E) and Marks (M)
1S SL(n) S CII 1S
2D SP S CII 1S;2D
3P NSP NS CI&II 1S;2D;3P
  • Respiratory stimulation by simultaneous action of type N substrates & succinate, with convergent electron flow in the NS-pathway for reconstitution of TCA cycle function.
  • OXPHOS capacity P (with saturating [ADP]), active OXPHOS state.
4Rot SP S CII 1S;2D;3P;4Rot
5Ama ROX 1S;2D;3P;4Rot;5Ama
  • Rox is the residual oxygen consumption in the ROX state, due to oxidative side reactions, estimated after addition of antimycin A (inhibitor of CIII). Rox is subtracted from oxygen flux as a baseline for all respiratory states, to obtain mitochondrial respiration (mt).
Step Respiratory state Pathway control ET-Complex Comment
## AsTm AsTmE CIV CIV
## Azd CHB


Questions.jpg


Click to expand or collaps

Strengths and limitations

+Reasonable duration of the experiment.
+This protocol can be extended with the Complex IV module.

Compare SUIT protocols