:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know that you can use the [[TIP2k-Module|Titration-Injection microPump (TIP2k)]] to perform an instrumental O<sub>2</sub> background on the [[O2k-sV-Module]]?''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know that you can use the [[TIP2k-Module|Titration-Injection microPump (TIP2k)]] to perform an instrumental O<sub>2</sub> background on the [[O2k-sV-Module]]?''' |
Revision as of 17:26, 16 March 2020
In discussion for page
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- (1) By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with the "BG Feedback "settings. - »O2k-sV-Module«
- (2) By adjusting the TIP2k setup with specific settings for the O2k-sV-Module, the instrumental O2 background can be performed in the O2k-sV-Module. - »O2k-sV-Module«
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?
- Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform an experimental troubleshooting on your own.-»Switch components approach«
- Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?
- See how to clean and dry the electrical connections of the OroboPOS-Connector.-»Cleaning the electrical connections«
- Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?
O2k-Open Support alert circulars
O2k-Open Support alert circular 2020(01): in prep
- New A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
- See how you can check that the two blue OroboPOS-Holders are intact. - »O2k-OroboPOS-Holder«
- New A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
- New Calibration parameters are transferred to experimental DatLab files
- Save calibration information from selected Marks while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - » Oxygen calibration - DatLab«
- Do you perform O2k-Chamber volume calibration?
- The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the O2k-sV-Module). - »O2k-Chamber MitoPedia«
- Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?
- New Calibration parameters are transferred to experimental DatLab files
- The O2 solubility factor varies according to the medium used and influences the air calibration. - »Oxygen solubility factor«